A Precise Gas Dilutor Based on Binary Weighted Critical Flows to Create NO2 Concentrations

Philipp Breitegger, Alexander Bergmann

    Publikation: Beitrag in Buch/Bericht/KonferenzbandBeitrag in einem KonferenzbandBegutachtung

    Abstract

    A gas diluter based on critical orifices was built and evaluated. The gas diluter is capable of creating dilution ratios of 1:1400 at a total flow of 6.5 L/min. An extended uncertainty analysis of gas concentrations and dilution ratios according to the Guide to the Expression of Uncertainty in Measurement was conducted. A gas cylinder of 5.16 ppm NO2 with a relative uncertainty of 1.5% (k = 1) can be diluted down to a concentration of 3.69 ppb NO2 (dilution ratio of 1:1400) at an uncertainty of 1.9% (k = 1). The results are in good agreement with reference NO2 measurements, conducted with a chemiluminescence detector (CLD, European reference method EN14211; 2005).
    Originalspracheenglisch
    TitelA Precise Gas Dilutor Based on Binary Weighted Critical Flows to Create NO2 Concentrations
    Seiten1-4
    Band2
    Auflage13
    DOIs
    PublikationsstatusVeröffentlicht - 7 Dez. 2018

    Publikationsreihe

    NameProceedings
    Herausgeber (Verlag)MDPI AG
    ISSN (Print)2504-3900

    Fingerprint

    Untersuchen Sie die Forschungsthemen von „A Precise Gas Dilutor Based on Binary Weighted Critical Flows to Create NO2 Concentrations“. Zusammen bilden sie einen einzigartigen Fingerprint.

    Dieses zitieren