Advanced design of MRI inversion pulses for inhomogeneous field conditions by optimal control

Christina Graf, Martin Söllradl, Christoph Stefan Aigner, Armin Rund, Rudolf Stollberger*

*Korrespondierende/r Autor/-in für diese Arbeit

Publikation: Beitrag in einer FachzeitschriftArtikelBegutachtung

Abstract

Non-selective inversion pulses find widespread use in MRI applications, where requirements on them are increasingly demanding. With the use of high and ultra-high field strength systems, robustness to (Formula presented.) and (Formula presented.) inhomogeneities, while tackling SAR and hardware limitations, has rapidly become important. In this work, we propose a time-optimal control framework for the optimization of (Formula presented.) - and (Formula presented.) -robust inversion pulses. Robustness is addressed by means of ensemble formulations, while allowing inclusion of hardware and energy limitations. The framework is flexible and performs excellently for various optimization goals. The optimization results are analyzed extensively in numerical experiments. Furthermore, they are validated, and compared with adiabatic RF pulses, in various phantom and in vivo measurements on a 3 T MRI system.

Originalspracheenglisch
Aufsatznummere4790
FachzeitschriftNMR in Biomedicine
Jahrgang35
Ausgabenummer11
DOIs
PublikationsstatusVeröffentlicht - 2022

ASJC Scopus subject areas

  • Molekularmedizin
  • Radiologie, Nuklearmedizin und Bildgebung
  • Spektroskopie

Fingerprint

Untersuchen Sie die Forschungsthemen von „Advanced design of MRI inversion pulses for inhomogeneous field conditions by optimal control“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren