Evaluating Characteristics of Electrostatic Discharge (ESD) Events in Wearable Medical Devices: Comparison with the IEC 61000-4-2 Standard

Mehdi Kohani, Aniket Bhandare, Li Guan, David Pommerenke, Michael G. Pecht*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

Electrostatic discharge (ESD) malfunctions in wearable medical devices have resulted in injuries and deaths. The U.S. Food and Drug Administration (FDA) recommends medical device manufacturers to qualify the electromagnetic compatibility of their products according to the IEC 60601-1-2 collateral standard, within which the IEC 61000-4-2 standard is the recommended ESD immunity test method. This paper shows that the IEC 61000-4-2 standard does not provide adequate immunity for wearable medical devices, due to the differences between the test setup and the real usage. The peak and maximum time derivative of the ESD current and transient magnetic fields were measured during realistic discharge scenarios, where a metal piece, in lieu of a wearable device, was worn around human hand or waist. At five voltage levels from 2 to 10 kV, the peak currents and their maximum time derivatives for realistic discharge scenarios were larger than the most severe discharges obtained from the IEC 61000-4-2 test configuration. This discrepancy between the current results from the IEC 61000-4-2 standard setup and the realistic discharge scenarios indicates that the ESD immunity standard for wearable devices needs to be changed with realistic test configurations, to prevent device malfunctions that can subsequently jeopardize patient safety.

Original languageEnglish
Pages (from-to)1304-1312
Number of pages9
JournalIEEE Transactions on Electromagnetic Compatibility
Volume60
Issue number5
DOIs
Publication statusPublished - 1 Oct 2018
Externally publishedYes

Keywords

  • Discharge current
  • electrostatic discharge (ESD)
  • medical device
  • transient magnetic field
  • wearable device

ASJC Scopus subject areas

  • Atomic and Molecular Physics, and Optics
  • Condensed Matter Physics
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Evaluating Characteristics of Electrostatic Discharge (ESD) Events in Wearable Medical Devices: Comparison with the IEC 61000-4-2 Standard'. Together they form a unique fingerprint.

Cite this